Recommended settings for easyPhaco

The Oertli easyPhaco technology is available in three different tip sizes. Incisions of sub–2-mm coaxial micro-phaco up to standard incision sizes of up to 3.2 mm can be used with easyPhaco.

Minimum   Venturi pump   Peristaltic pump / SPEEP pump  
    Vacuum (mmHg) Vacuum effect (%) Flow (ml/min) Vacuum (mmHg)
easyTip CO-MICS
for incisions of 1.6-1.8 mm
Phaco 1 (Grooving)
Phaco 2
80
250
60
60
18
20
50
250
easyTip 2.2 mm
for incisions of  2.2-2.4 mm
Phaco 1 (Grooving)
Phaco 2
80
400
60
60
18
40
50
400
easyTip 2.8 mm
for incisions of 2.8-3.2 mm
Phaco 1 (Grooving)
Phaco 2
80
400
60
60
18
40
50
400
           

--> Recommended bottle height: 100 cm above the patient eye
--> Recommended additional settings: Pulse mode, 40 Hz, 60% cooling for Phaco 2.

easyPhaco is a development from Oertli R&D in scientific collaboration with Professor Rupert Menapace, Vienna.

Please contact your sales partner to determine your optimal easyPhaco values.